Apparatus:OBP-30KVOBP-50KV System

Technical Specifications & Test Capabilities

Parameter / MetricSpecification Range & AccuracyOperational Notes
Poling Voltage RangeDC 0 to 10 kV / 30 kV / 50 kVProgrammable ramp rate ≤ 1000 V/s
Temperature RangeRoom Temperature to 180°CInsulating silicone oil bath heating
Safety & ProtectionHardware current limit 1 mA with arc suppressionPrevents sample cracking or electrical breakdown
Process ModesRoom-temp poling, Constant-temp poling, Synchronous cooling polingCustom poling cycles tailored to Curie point

Specimen Geometry & Preparation Guide

Material FormRecommended DimensionsDimensional LimitsElectrode & Poling Notes
Bulk Ceramic Discs & PlatesDiameter 20 mm, thickness 1–2 mmDiameter 2–50 mm, thickness ≤ 12 mmStrictly zero electrode paste on side walls to prevent arc flashover
Thin Films on Substrates2D dimension ≤ 40 mm, thickness < 5 mmDouble-sided electrodedCompatible with high-voltage immersion fixtures

Testing Function

As-sintered ferroelectric ceramics possess randomly oriented spontaneous polarization domains, resulting in zero net macroscopic piezoelectric response. High-voltage oil bath poling aligns internal domains along an intense external DC electric field (up to 50 kV) under elevated temperatures (up to 180°C) within insulating dielectric oil, imparting permanent piezoelectric properties without air breakdown.

Testing Conditions

Primary Instruments: OBP-30KV, OBP-50KV System
Standard Turnaround: Within 5 working days
Key Standards: GB/T 3389-2008 Methods for Piezoelectric Ceramic Material Poling Protocols

Sample Requirements

Please review the detailed specimen dimensions, electrode requirements, and poling conditions in the table above. Custom dimensions and special sample geometry can be accommodated upon technical consultation.

Testing Standards

  • GB/T 3389-2008 Methods for Piezoelectric Ceramic Material Poling Protocols

Test Results

Clients receive comprehensive test reports with raw exported ASCII/txt data files, high-resolution curve plots (temperature spectra, impedance plots, or hysteresis loops), and experimental condition parameters stamped by our CMA accredited testing center.