Quasi-Static Piezoelectric d33 / d31 / d15 Testing
CMA Certified Lab (No. 261721340056)⏱ Within 3 working days

Quasi-Static Piezoelectric d33 / d31 / d15 Testing

Apparatus:PEAL-1000

Technical Specifications & Test Capabilities

Parameter / MetricSpecification Range & AccuracyOperational Notes
Measurement Range0 to 6,000 pC/NDirect traceability method
Static Force Range1 N to 10 N (Resolution: 0.1 N)Calibrated pneumatic/mechanical clamp
Dynamic Force Range0.05 N to 0.5 N (Resolution: 0.05 N)Traceable low-noise mechanical excitation
Measurement ModesLongitudinal d33, Transverse d31, Shear d15Supports cylinders, discs, square plates, and films

Specimen Geometry & Preparation Guide

Material FormRecommended DimensionsDimensional LimitsElectrode & Poling Notes
Piezoelectric Ceramic Discs/CylindersDiameter 15–20 mm, thickness 1–2 mmDiameter ≤ 35 mm, thickness ≤ 25 mmElectroded and poled; aged ≥ 24 hours prior to measurement
Transverse d31 Bar Specimen8 × 5 × 3 mmLength 10–35 mm, thickness 2–10 mmElectrodes on major faces, poled along thickness
Shear d15 Plate Specimen15 × 5 × 1 mmThickness ≥ 3 mm, 2D ≤ 25 mmSpecialized poling and re-metallization protocol
Piezoelectric Films & PolymersDiameter 10 mm, thickness ≥ 10 µmThickness ≥ 10 µm, 2D ≤ 10 mmPoled bare film (no metal electrode required)

Testing Function

The piezoelectric coefficient d33 defines the electromechanical coupling strength between mechanical force and electric displacement. Utilizing the direct traceability method, our calibrated PEAL-1000 system applies dynamic low-frequency mechanical stress and measures the generated charge with precision electrometry. We provide accredited characterization of d33 (longitudinal), d31 (transverse), and d15 (shear) across ceramics, single crystals, polymers, and electrospun membranes.

Testing Conditions

Primary Instruments: PEAL-1000
Standard Turnaround: Within 3 working days
Key Standards: GB/T 11309-1989 Test Methods for Quasi-Static Testing of Longitudinal Piezoelectric Strain Constant d33; SJ/T 11469-2014 Electronic Industry Standard of China: Shear Piezoelectric Constant d15; IEEE 176 Standards on Piezoelectricity

Sample Requirements

Please review the detailed specimen dimensions, electrode requirements, and poling conditions in the table above. Custom dimensions and special sample geometry can be accommodated upon technical consultation.

Testing Standards

  • GB/T 11309-1989 Test Methods for Quasi-Static Testing of Longitudinal Piezoelectric Strain Constant d33
  • SJ/T 11469-2014 Electronic Industry Standard of China: Shear Piezoelectric Constant d15
  • IEEE 176 Standards on Piezoelectricity

Test Results

Clients receive comprehensive test reports with raw exported ASCII/txt data files, high-resolution curve plots (temperature spectra, impedance plots, or hysteresis loops), and experimental condition parameters stamped by our CMA accredited testing center.