Non-Contact Corona Poling Service for Thin Films (±20 kV)
CMA Certified Lab (No. 261721340056)⏱ Within 5 working days

Non-Contact Corona Poling Service for Thin Films (±20 kV)

Apparatus:SPD Corona Poling SystemFCP Film Corona Polarizer

Technical Specifications & Test Capabilities

Parameter / MetricSpecification Range & AccuracyOperational Notes
Corona Voltage RangeDC 0 to 10 kV / 20 kV / -20 kV (Positive / Negative)Single-needle or multi-needle needle-mesh array
Temperature RangeRoom Temperature to 180°CUniform heated sample ground stage
Grid StabilizationBias-controlled grid mesh for uniform surface potentialEliminates localized pinhole breakdown
Target SubstratesPVDF, P(VDF-TrFE), cellular polypropylene, electrets, powdersSpecimen dimensions up to A5 (20 × 20 cm)

Specimen Geometry & Preparation Guide

Material FormRecommended DimensionsDimensional LimitsElectrode & Poling Notes
Thin Film CouponsDiameter ≤ 45 mm, single-sided bottom electrodeThickness 5–200 µmTop surface remains bare for charge deposition
Large-Area Polymer SheetsA5 size or 20 × 20 cmSingle-sided metallized bottomUniform area poling
Electret & Piezoelectric Powders≤ 2 g batchDeposited/fixed onto conductive metallic plateAllows high-field dipole alignment without pelletizing

Testing Function

Conventional contact poling of thin polymer films or porous cellular membranes frequently suffers from pinhole dielectric breakdown caused by microscopic defects. Corona poling employs sharp needle-tip gas ionization and grid control to deposit charged ions gently onto the exposed dielectric surface, generating an immense electric field that aligns dipoles without destructive electrical punch-through.

Testing Conditions

Primary Instruments: SPD Corona Poling System, FCP Film Corona Polarizer
Standard Turnaround: Within 5 working days
Key Standards: IEEE 176 / ASTM Standards on Corona Charging and Electret Stability

Sample Requirements

Please review the detailed specimen dimensions, electrode requirements, and poling conditions in the table above. Custom dimensions and special sample geometry can be accommodated upon technical consultation.

Testing Standards

  • IEEE 176 / ASTM Standards on Corona Charging and Electret Stability

Test Results

Clients receive comprehensive test reports with raw exported ASCII/txt data files, high-resolution curve plots (temperature spectra, impedance plots, or hysteresis loops), and experimental condition parameters stamped by our CMA accredited testing center.