Microwave Dielectric Cavity Testing (1 GHz to 40 GHz)
CMA Certified Lab (No. 261721340056)⏱ Within 5 working days

Microwave Dielectric Cavity Testing (1 GHz to 40 GHz)

Apparatus:Keysight E5071C Network AnalyzerTE01δ Hakki-Coleman Closed Cavity

Technical Specifications & Test Capabilities

Parameter / MetricSpecification Range & AccuracyOperational Notes
Testing Frequency2 GHz to 19 GHzDependent on specimen geometry and permittivity
Relative Permittivity Rangeεr ≥ 1.5 (Accuracy: |Δεr| ≤ 0.2% εr)Electromagnetic finite element modal analysis
Dielectric Loss Rangetan δ ≤ 10⁻² (Accuracy: |Δtanδ| ≤ 2% tanδ + 1.5×10⁻⁶)Resolves loss tangents down to 10⁻⁵
Derived Figures of MeritUnloaded Q0, Qf factor (GHz), Resonance frequency f0Essential for 5G/6G RF substrate screening

Specimen Geometry & Preparation Guide

Material FormRecommended DimensionsDimensional LimitsElectrode & Poling Notes
Precision Cylindrical RodDiameter D ≈ 10 mm, Height H ≈ 5 mmD = 6–15 mm, H = 3–7 mm (Tolerance ≤ 0.2%)Parallel polished planar ends, cylindrical symmetry without chips

Testing Function

The TE01δ closed cylindrical metallic cavity resonator method (Hakki-Coleman configuration) represents the international gold standard for measuring the microwave permittivity and dielectric loss of low-loss RF ceramics and microwave dielectric substrates. Coupled to a Keysight E5071C vector network analyzer with rigorous electromagnetic finite-element analysis, our laboratory captures resonant peak f0 and unloaded quality factor Q0 to extract true εr, tan δ, and the signature Qf quality factor.

Testing Conditions

Primary Instruments: Keysight E5071C Network Analyzer, TE01δ Hakki-Coleman Closed Cavity
Standard Turnaround: Within 5 working days
Key Standards: IEC 61338-1-3 Waveguide Type Dielectric Resonators - Measurement of Dielectric Properties; IPC-TM-650 Test Methods for Stripline and Resonant Cavity High-Frequency Permittivity

Sample Requirements

Please review the detailed specimen dimensions, electrode requirements, and poling conditions in the table above. Custom dimensions and special sample geometry can be accommodated upon technical consultation.

Testing Standards

  • IEC 61338-1-3 Waveguide Type Dielectric Resonators - Measurement of Dielectric Properties
  • IPC-TM-650 Test Methods for Stripline and Resonant Cavity High-Frequency Permittivity

Test Results

Clients receive comprehensive test reports with raw exported ASCII/txt data files, high-resolution curve plots (temperature spectra, impedance plots, or hysteresis loops), and experimental condition parameters stamped by our CMA accredited testing center.

Dual-Funnel Validation

Turnkey Equipment for Your Own Laboratory

Prefer in-house characterization? Every test in our laboratory is performed using MatMeas turnkey instruments. Verify measurement quality with sample testing before capital procurement.

Request High-Precision Testing Solutions

Get expert guidance and customized instruments for your functional materials project.